Screening Services and
Training
Last Update 10/06/2005
25092 Natama
Ct
Based upon a procedure written by Screening Systems, Inc.
Laguna Hills, CA 92653-5607
How to: Lower the Chamber Operating Range to below
–60 degrees C on SSCTM-2000 Systems
This procedure applies to systems manufactured by Screening Systems,
Inc. that use the model QRSTM-2000 vibration controllers.
The temperature low limit for some model QRSTM systems is set to
-60° C. To lower the limit below this value on
systems
controlled by the SSC-2000 controller requires a few steps be
accomplished to assure proper system operation and integrity.
The first step is to lower the Chamber Absolute Temperature setting in
the System Configuration. From the main menu across the top
of
the screen, select “Configure” then
“System”
from the resulting pull-down menu. The following window will
appear.
Change the circled parameter to the lowest temperature to be set or
programmed during the test. The lowest value that can be set
is
–100 degrees C. To avoid errors, however, it is
recommended
that the setting not be any lower than necessary for a particular test.
After selecting “EXIT (F1)” and saving the changes,
adjust
the system’s independent Low Temp Limit device located on the
front panel of most systems.
The limit detector shown above, Honeywell UDC 2000, has two adjustments
that must be changed. The first limit is adjusted by changing
the
“Setpoint” value 10 to 15 degrees lower than the
lowest
value that will be programmed. When this limit is exceeded
the
redundant LN2 valve is de-energized until the detector is reset.
To change the “setpoint”, push the
“Display”
button on the Low Temp Limit device and use the down and up arrow
buttons to select the desired value.
The second limit within this device alerts the computer controller that
this temperature limit has been exceeded.
This action results in an audible warning along with a red visual
warning on the controller screen indicating that an external
temperature limit has been exceeded. In addition, the
controller
command to the LN2 control valve is set to zero so no more cooling is
called for. This second limit is usually set between 5 to 10
degrees lower than the lowest value that will be programmed.
To change this limit, push the “Set Up” button
until
“ALARM” shows in the display. Then push
the
“Function” button until
“A1S1VA” is shown in
the display. Using the up and down arrows, adjust the setting
to
the desired value. To enter the new value and return to the
top
display, push the “DISPLAY” button once.
To leave the
value unchanged, push the “RESET” button once and
then the
“DISPLAY” button.
An alternate limit detector, RKC Instruments, Inc. Model CB900L has a
single limit that both turns off the LN2 valve and alerts the computer
controller when the low limit is exceeded.
To change the low limit on this device starting from the PV/SV display,
push the “SET” button once. Press the
“left
arrow” button until the desired integer is highlighted then
push
the up and down arrows to change the integer. To accept the
new
setting and return to the PV/SV display, push the
“SET”
button 3 times.
The final step in the procedure is to install or verify the
installation of a thick table mat. The purpose of the mat is
to
protect the table from the extreme cold and to decrease the LN2
consumed during a test. An additional benefit of the mat
after a
cold soak is to decrease the amount of warm-up time required before the
doors can be opened without causing condensation to form on the table
surface.
This completes the procedure.
For technical assistance contact Screening Services and Training via
e-mail. scrserv@cox.net