Screening Services and
Training
Last Update 10/06/2005
25092 Natama
Ct
Based upon a procedure written by Screening Systems, Inc.
Laguna Hills, CA 92653-5607
How to Adjust the SSCTM-2000 Controller Null Pacing, RampGuard, and
SoakGuard Parameters.
Null Pacing
The Null Pacing capability of the SSCTM –2000 Controller
monitors
the differences between the Profile Temperature Command and the reading
of the temperature device selected as the Temperature Feedback.
When the difference between these two values exceeds the
operator-selected temperature difference limit of the SoakGuard or the
RampGuard, the temperature Profile Generator holds at the present
level, Vibration (if actively running) turns off, and the Null Pacing
Mode Timeout timer starts counting. This is called the Null
Pacing Mode.
By holding these commands, the Profile Generator gives the Feedback
temperature time to “catch-up” to the Command thus
decreasing the difference between the values. When the
difference
between these values decreases to 50% of the operator-selected
temperature difference, SoakGuard or RampGuard settings, the Generator
resumes the profile.
Null Pacing may be configured to automatically terminate a temperature
profile after being continuously in the Null-Pacing Mode longer than
the operator-selected Null Pacing Mode Timeout setting.
Selecting
OFF will disable this Timeout feature.
SoakGuard
Located in the bottom row of the Program input pop-up window, this
control allows the user to set the deviation between the Setpoint and
the Feedback temperature that will cause the SSC™-2000 to
enter
the Null-Pacing Mode when in a Soak segment. Any segment with
identical starting and ending temperature Setpoint values is a Soak
segment. Selecting OFF in this control will disable the
Null-Pacing Mode during all Soak segments.
In the figure below, the Soak guard is configured as 5 degrees, which
means that the SSC™-2000 will enter the Null-Pacing Mode in
any
Soak segment whenever the Feedback temperature is more than 5 degrees
above or below the current temperature Setpoint Command value.
RampGuard
Also located in the bottom row of the Program input pop-up window, this
control allows the user to set the deviation between the Setpoint and
the Feedback temperature that will cause the SSC™-2000 to
enter
the Null-Pacing Mode when in a Ramp segment. Any segment with
different starting and ending temperature Setpoint values is a Ramp
segment. Selecting OFF in this control will disable the
Null-Pacing Mode during all Ramp segments.
Also in the figure, the Ramp guard is configured as 15 degrees, which
means that the SSC™-2000 will enter the Null-Pacing Mode in
any
Ramp segment whenever the Feedback temperature is more than 15 degrees
above or below the current temperature Setpoint Command value.
For technical assistance contact Screening Services and Training via
e-mail. scrserv@cox.net